• Microsurgery · Oct 2011

    Optical requirements on magnification systems for intracranial video microsurgery.

    • Florian H Ebner, Jakob S Marquardt, Bernhard Hirt, Marcos Tatagiba, and Frank Duffner.
    • Department of Neurosurgery, Eberhard-Karls-University, Tuebingen Institute of Clinical Anatomy, Tuebingen.
    • Microsurgery. 2011 Oct 1;31(7):559-63.

    IntroductionThe basic idea of video-microsurgery is the improvement of ergonomic conditions in microsurgical procedures by replacing the bulky operating microscope with a compact videosystem.ObjectiveTo specify optical requirements on a videosystem for microsurgical intracranial procedures in neurosurgery.MethodsDuring 27 microsurgical intracranial procedures (12 cerebellopontine angle and 15 supratentorial) zoom factor, focus distance and illumination parameters of the operating microscope were continuously recorded. Ergonomic aspects were documented as well.ResultsThe zoom factor ranged from 1.7 to 13.5 in CPA procedures and from 1.4 to 13.4 in supratentorial procedures. The focus distance ranged from 180 mm to 367 mm in CPA procedures and from 188 mm-472 mm in supratentorial procedures.ConclusionFrom an optical point of view current operating microscopes meet the requirements of intracranial microneurosurgery. However, ergonomically further developments are highly desirable. Video microsurgery is a promising field and could hold a solution to this problem.Copyright © 2011 Wiley-Liss, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        

    hide…