• Anesthesia and analgesia · Feb 2005

    Comparative Study

    The ProSeal has a shorter life-span than the Classic laryngeal mask airway.

    • Sarah Doneley, Joseph Brimacombe, Christian Keller, and Achim von Goedecke.
    • Department of Anaesthesia, Cairns Base Hospital, The Esplanade, Cairns 4870, Australia.
    • Anesth. Analg. 2005 Feb 1;100(2):590-3.

    AbstractWe tested the hypothesis that the life-span of the ProSeal laryngeal mask airway (LMA) is shorter than the Classic LMA. We also compared residual cuff volumes and changes in elastance and permeability with use. Six new size 4 ProSeal LMAs and 6 new size 4 Classic LMAs were tested. Each LMA underwent a simulated clinical use cycle until it failed the pre-use check tests. The simulated clinical use cycle comprised: i) inflating the cuff to the maximum recommended volume for 1 h; ii) immersion in an enzymatic solution for 3 min; iii) washing the external and internal surfaces; iv) automatic washing for 14 min (< or =85 degrees C); v) drying for 30 min (< or =75 degrees C); and vi) autoclaving at 134 degrees C for 4 min at 206 kPa. Before the first cycle and every 10 subsequent cycles, the cuff was inflated with 40 mL air and the intracuff pressure was measured immediately and 3 h later. The initial intracuff pressure was taken to be an inverse measure of the elastance or resistance to deformation, and the intracuff pressure change were taken as a measure of the permeability. The residual cuff volume was determined for 10 Classic and 10 ProSeal size 4 LMAs using a gas dilution technique. The mean +/- sd (range) longevity for the ProSeal LMA and Classic LMA was 82 +/- 23 (45-109) uses and 133 +/- 35 (76-176) uses, respectively. The ProSeal LMA has a shorter life-span than the Classic LMA (P = 0.01). For the ProSeal LMA, there was no change in elastance or permeability with use. For the Classic LMA, there was a decrease in elastance (P < 0.0001) and an increase in permeability (P < 0.0001) with use. The residual cuff volume was higher for the ProSeal LMA (2.6 +/- 1.3 mL versus 1.5 +/- 0.9, P = 0.04). We conclude that the life-span of the ProSeal LMA is shorter than the Classic LMA, but both exceed the manufacturer's recommendations of 40 uses. We recommend that reusable LMA devices be discarded when they fail the pre-use check tests, rather than after a specific number of uses.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…