Accuracy and precision of three different methods to determine Pco2 (Paco2 vs. Petco2 vs. Ptcco2) during interhospital ground transport of critically ill and ventilated adults.
- Jochen Hinkelbein, Florian Floss, Christof Denz, and Heiner Krieter.
- Clinic for Anesthesiology and Intensive Care Medicine, Ruprecht-Karls-University Heidelberg, University Hospital Mannheim, Mannheim, Germany. jochen.hinkelbein@anaes.ma.uni-heidelberg.de
- J Trauma. 2008 Jul 1;65(1):10-8.
BackgroundInterhospital transportation of critically ill and mechanically ventilated patients represents a common, yet difficult problem. Three different methods to determine Pco2 during transport are available: arterial blood gas analysis (Paco2), end-tidal (Petco2) and transcutaneous (Ptcco2) measurement. The aim of the present study is to analyze accuracy and precision of those different methods simultaneously in critically ill and ventilated adults during interhospital transport.MethodsPatients scheduled for interhospital transport were investigated after approval of the local ethics committee in the prospective study. Pco2 was determined five times in each patient during the transport simultaneously by (1) arterial blood gas analysis (Paco2[Immediate Response Mobile Analyzer, IRMA]), (2) end-tidal (Petco2), and (3) transcutaneous (Ptcco2) measurements. The results were compared with an in-hospital reference measurement performed by an ABL 625 blood gas analyzer (Paco2[ABL625]). For statistical analysis the Bland-Altman method was used. A p < 0.05 was considered statistically significant.ResultsOne hundred seventy data sets (Paco2[IRMA], Paco2[ABL625], Petco2, Ptcco2) were obtained in 34 patients (61 years +/- 16 years old; 19 male patients, 15 female patients). The mean Paco2(ABL625) was 43.2 mm Hg +/- 8.8 mm Hg ranging from 24.9 mm Hg to 72.4 mm Hg. Bland-Altman analysis revealed a bias and precision of -0.6 mm Hg +/- 2.5 mm Hg for the arterial blood gas analysis with the mobile IRMA device and -0.6 mm Hg +/- 7.5 mm Hg for the transcutaneous measurement (p > 0.05). Bias and precision (-5.3 mm Hg +/- 6.1 mm Hg) of endexpiratory CO2-measurement differed significantly (p < 0.003) when compared with the reference.ConclusionsDuring interhospital transport Paco2(IRMA) and Ptcco2 provide the best accuracy when compared with the reference measurement. Patients who either require a tight control of Pco2 or endured lengthy transportation could benefit greatly from the combination of expiratory capnography with mobile arterial blood gas analysis or the transcutaneous measurement of Pco2.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.