Anesthesia and analgesia · Oct 2010
Comparative StudyTwo serial check valves can prevent cross-contamination through intravenous tubing during total intravenous anesthesia.
- Oliver C Radke, Katrin Werth, Margarete Borg-von-Zepelin, Petra Saur, and Christian C Apfel.
- Klinik und Poliklinik für Anästhesiologie und Intensivtherapie, Dresden, Germany. oliver.radke@uniklinikum-dresden.de
- Anesth. Analg. 2010 Oct 1;111(4):925-8.
BackgroundNonsterile handling of propofol for anesthesia has been linked with severe sepsis and death. Placing a single check valve in the IV tubing does not prevent retrograde ascension of pathogens into propofol-filled syringes, so we designed an IV tubing set with multiple check valves. To estimate the efficacy of this design, we measured the concentration of pathogens detected upstream in the IV tubing in relation to the pathogen concentration in a model of a contaminated patient.MethodsA glass container with a rubber sealed port was filled with a suspension of either bacteria or phagocytes and kept at 37°C ("contaminated patient" model). A bag of normal saline was connected to an IV cannula, punctured through the rubber sealed port of the patient model. Two additional sidestream infusion lines were connected to syringes in 2 standard infusion pumps. One of the syringes contained propofol and the other contained normal saline as a substitute for an opioid preparation. After 5 hours of infusion, we obtained samples from different parts of the infusion lines and syringes. The samples were streaked out on blood agar plates and incubated at 37°C for 24 hours. We repeated this experiment with 6 different pathogens.ResultsWe incubated 825 agar plates. Whereas the concentration of bacteria and phagocytes in the "patient" had significantly increased during the 5-hour experiments (positive control), no bacterial growth could be detected in any of the incubated plates.ConclusionThe data from this experimental setting suggest that the design with multiple check valves in paired configuration prevents retrograde contamination. Of note, this does not permit the reuse of propofol syringes because reusing is against the manufacturer's recommendations.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.