• Acta Neurochir. Suppl. · Jan 2012

    Comparative Study

    Experimental comparison of the measurement accuracy of the Licox(®) and Raumedic (®) Neurovent-PTO brain tissue oxygen monitors.

    • Matthias H Morgalla, R Haas, G Grözinger, Christian Thiel, Karolin Thiel, Martin U Schuhmann, and Martin Schenk.
    • Department of Neurosurgery, Eberhard Karls University Hospital, Tübingen, Germany. matthias.morgalla@med.uni-tuebingen.de
    • Acta Neurochir. Suppl. 2012 Jan 1;114:169-72.

    BackgroundOnly a few experimental reports are available on the direct comparison of Licox(®) and Raumedic(®)-Neurovent-PTO brain tissue oxygen pressure (P(br)O(2)) monitors. We compared the two systems regarding their measurement properties under experimental in vitro and in vivo conditions.Materials And MethodsEight Licox(®) and Raumedic(®) Neurovent-PTO(®) sensors were tested for 10 min at 37°C, atmospheric pressure, at an oxygen content of 0% and 100% before and after the in vivo test. The same probes were implanted in German landrace pigs, which underwent hepatectomy. The mean P(br)O(2) values were recorded every minute. An O(2) challenge with inhalation of 100% O(2) for 10 min was performed 2 h post-abdominal surgery.ResultsAt 0% O(2) content values varied from 0.2 to 7 mmHg, at 100% O(2) content from 130 to 165 mmHg. No difference between probes was found. In vivo tests: Raumedic® showed higher P(br)O(2) values (mean +6.3 mmHg, p < 0.0001) compared with Licox®. During O(2) challenge, both probes responded similarly; however, Raumedic(®) had a 10% higher response amplitude (p < 0.005). After explantation there was again no difference between the two sensors.ConclusionRaumedic(®) sensors measured higher P(br)O(2) values. There was no significant difference regarding overall measurement of in vitro accuracy between the two probes, which proved to be robust when used consecutively for longer periods and in different environments.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…