• J Neurosurg Spine · Nov 2010

    Comparative Study

    Clinical accuracy of three-dimensional fluoroscopy-based computer-assisted cervical pedicle screw placement: a retrospective comparative study of conventional versus computer-assisted cervical pedicle screw placement.

    • Yoshimoto Ishikawa, Tokumi Kanemura, Go Yoshida, Zenya Ito, Akio Muramoto, and Shuichiro Ohno.
    • Department of Spine and Orthopaedic Surgery, Konan Kosei Hospital, Konan City, Japan. chita.peninsula@gmail.com
    • J Neurosurg Spine. 2010 Nov 1;13(5):606-11.

    ObjectThe authors performed a retrospective clinical study to evaluate the feasibility and accuracy of cervical pedicle screw (CPS) placement using 3D fluoroscopy-based navigation (3D FN).MethodsThe study involved 62 consecutive patients undergoing posterior stabilization of the cervical spine between 2003 and 2008. Thirty patients (126 screws) were treated using conventional techniques (CVTs) with a lateral fluoroscopic view, whereas 32 patients (150 screws) were treated using 3D FN. Screw positions were classified into 4 grades based on the pedicle wall perforations observed on postoperative CT.ResultsThe prevalence of perforations in the CVT group was 27% (34 screws): 92 (73.0%), 12 (9.5%), 6 (4.8%), and 16 (12.7%) for Grade 0 (no perforation), Grade 1 (perforation < 1 mm), Grade 2 (perforation ≥ 1 and < 2 mm), and Grade 3 (perforation ≥ 2 mm), respectively. In the 3D FN group, the prevalence of perforations was 18.7% (28 screws): 122 (81.3%), 17 (11.3%), 6 (4%), and 5 (3.3%) for Grades 0, 1, 2, and 3, respectively. Statistical analysis showed no significant difference in the prevalence of Grade 1 or higher perforations between the CVT and 3D FN groups. A higher prevalence of malpositioned CPSs was seen in Grade 2 or higher (17.5% vs 7.3%, p < 0.05) in the 3D FN group and Grade 3 (12.7% vs 7.3%, p < 0.05) perforations in the CVT group. The ORs for CPS malpositioning in the CVT group were 2.72 (95% CI 1.16-6.39) in Grade 2 or higher perforations and 3.89 (95% CI 1.26-12.02) in Grade 3 perforations.ConclusionsThree-dimensional fluoroscopy-based navigation can improve the accuracy of CPS insertion; however, severe CPS malpositioning that causes injury to the vertebral artery or neurological complications can occur even with 3D FN. Advanced techniques for the insertion of CPSs and the use of modified insertion devices can reduce the risk of a malpositioned CPS and provide increased safety.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…