Technol Health Care · Jan 2008
Interference of neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators: an in vitro study.
- Salome Ryf, Thomas Wolber, Firat Duru, and Roger Luechinger.
- Institute for Biomedical Engineering, University and ETH Zurich, Switzerland.
- Technol Health Care. 2008 Jan 1;16(1):13-8.
AbstractPermanent magnets may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets have become widely available in recent years and are incorporated in various articles of daily life. We conducted an in-vitro study to evaluate the ability of NdFeB magnets for home and office use to cause interference with cardiac pacemakers and ICDs. The magnetic fields of ten NdFeB magnets of different size and shape were measured at increasing distances beginning from the surface until a field-strength (B-field) value of 0.5 mT was reached. Furthermore, for each magnet the distance was determined at which a sample pacemaker switched from magnet mode to normal mode. Depending on the size and remanence of individual magnets, a B-field value of 0.5 mT was found at distances ranging from 1.5 cm to 30 cm and a value of 1 mT at distances from 1 cm to 22 cm. The pacemaker behavior was influenced at distances from 1 cm to 24 cm. NdFeB magnets for home and office use may cause interference with cardiac pacemakers and ICDs at distances up to 24 centimeters. Patient education and product declarations should include information about the risk associated with these magnets.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.