    [Recirculation of urea and dialysis efficiency using dual-lumen dialysis catheters in various locations: may the venous lumen of the catheter be used as the arterial lumen and vice versa?].

    • Sinisa Sefer, Petar Kes, Vesna Degoricija, Branko Heinrich, and Mislav Vrsalović.
    • KB Sestre milosrdnice, Zavod za nefrologiju i dijalizu, Vinogradska c. 29, 10 000 Zagreb.
    • Lijec Vjesn. 2003 Jan 1;125(1-2):1-5.

    AimTo measure percentage of urea recirculation in hemodialysis by a dual-lumen central venous catheters of various localisations (e.g. jugular, subclavian and femoral), and also in cases when the venous lumen of the catheter is used as arterial lumen, and vice versa, and under these conditions to evaluate the efficiency of hemodialysis based on the measurements of urea reduction ratio.MethodsPercentage of urea recirculation was measured in the patients with acute and chronic renal failure who underwent hemodialysis by a temporary dual-lumen central venous catheter of different localisations (jugular n = 16, subclavian n = 20, femoral n = 20). The measurements were done in cases when arterial line was connected to arterial lumen of the catheter, and venous line to the venous lumen of the catheter (correct connection = RI), and vice versa, i.e., when arterial line was connected to venous lumen of the catheter, and venous line was connected to arterial lumen of the catheter (reversed connection = R2). The efficiency of hemodialysis was evaluated by measuring urea reduction ratio in cases with the reversed (R2) connection.ResultsStatistically significant differences were found between R1 and R2 for jugular catheters (R1 = 2.38 +/- 1.09%, R2 = 7.59 +/- 1.42%, n = 16, p = 0), for subclavian catheters (R1 = 3.03 +/- 3.15%, R2 = 15.8 +/- 7.18, n = 20, p = 0), and for femoral catheters (R1 = 9 +/- 6.56%, R2 = 29.2 +/- 11.8%, n = 20, p = 0). Statistically significant differences were also found between R1 of jugular catheters and R1 of subclavian catheters (2.38 +/- 1.09%, n = 16 and 3.03 +/- 3.15%, n = 20, p = 0.0001), further on between R1 of jugular and R1 of femoral catheters (2.38 +/- 1.09%, n = 16 and 9 +/- 6.56%, n = 20, p = 0), as well as between R1 of subclavian and R1 of femoral catheters (2.38 +/- 1.09%, n = 20 and 9 +/- 6.56%, n = 20, p = 0.0001). Also statistically significant differences were found between R2 of jugular and subclavian catheters (7.59 +/- 1.42%, n = 16 and 15.8 +/- 7.18%, n = 20, p = 0.0003), between R2 of jugular and femoral catheters (7.59 +/- 1.42%, n = 16 and 29.2 +/- 11.8%, n = 20, p = 0.0007), and between R2 of subclavian and R2 of femoral catheters (15.8 +/- 7.18%, n = 20 and 29.2 +/- 11.8%, n = 20, p = 0.0029). The measurements of urea reduction ratio that we have done for some of the catheters under R2 conditions showed a statistically significant difference between femoral and subclavian catheters (51.45 +/- 5.62%, n = 20 and 63.75 +/- 7.61, n = 20, p = 0), and between femoral and jugular catheters (51.45 +/- 5.62%, n = 20 and 64.3 +/- 5.23%, n = 16, p = 0). No statistical differences were found in urea reduction ratio between jugular and subclavian catheters (64.3 +/- 5.23, n = 16 and 63.75 +/- 7.61%, n = 20, p = 0.8).ConclusionWhen hemodialysis is delivered by a correct blood lines connection the measurements have shown a decrease in urea recirculation by 5% for jugular and subclavian catheters, and 5%-10% for femoral catheters. However, in cases when the venous lumen of the catheter is used as an arterial lumen, and vice versa, urea recirculation is below 10% for jugular catheters; whereas in femoral catheters the percentage is higher than 20%. Despite so high percentage of urea recirculation, that we obtained for such use of the catheters, urea reduction ratio in hemodialysis via jugular and subclavian catheters is > 60%, whereas via femoral catheters, the percentage is significantly lower. Consequently, the efficiency of hemodialysis is reduced, and such use of femoral catheters should therefore be avoided.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.