Randomized Controlled Trial
Parametric digital subtraction angiography imaging for the objective grading of collateral flow in acute middle cerebral artery occlusion.
- Wan-Ling Wen, Yi-Bin Fang, Peng-Fei Yang, Yong-Wei Zhang, Yi-Na Wu, Hui Shen, Jia-Jia Ge, Yi Xu, Bo Hong, Qing-Hai Huang, and Jian-Min Liu.
- Department of Neurosurgery, Changhai Hospital, Shanghai, China.
- World Neurosurg. 2016 Apr 1; 88: 119-125.
PurposeTo report the feasibility of parametric color-coded digital subtraction angiography (DSA) in complementing the traditional, subjective way of leptomeningeal collateral assessment in acute middle cerebral artery (MCA) occlusions.MethodsThirty-three consecutive patients with acute MCA occlusion who received endovascular treatment were recruited for investigation. Eighteen of 33 consecutive patients were included. The target downstream territory (TDT) of MCA and reference point at terminal internal carotid artery of each patient was contoured by 5 raters independently on the basis of anteroposterior 2-dimensional DSA. Two parameters of relative maximum density of TDT (rDensitymax) and peak time interval (ΔPT) between reference and TDT were extracted by the use of parametric DSA analysis software. Interrater reliability was tested with intraclass correlation coefficients. Parameters with sufficient interrater reliability entered validity evaluation. Then, the correlation test with the American Society of Interventional and Therapeutic Neuroradiology collateral grading system and efficacy in predicting favorable clinical outcome was evaluated.ResultsThe intraclass correlation coefficient of rDensitymax and ΔPT were 0.983, 95% confidence interval 0.968-0.993 and 0.831, 95% confidence interval 0.705-0.923, respectively. The parameter rDensitymax showed a strong correlation with the American Society of Interventional and Therapeutic Neuroradiology collateral grading system score (r of Spearman correlation test = 0.869, P < 0.001) and mRS at 3 months (partial correlation coefficient = 0.616, P = 0.009), whereas ΔPT_average did not. A cut-off point of 0.224 in rDensitymax predicted a favorable clinical outcome with high sensitivity and specificity.ConclusionsThe relative maximum contrast density of MCA territory on 2-dimensional DSA measured by parametric imaging technique appears to be a simple and reliable metric for the assessment of leptomeningeal collaterals in cases of acute MCA occlusion.Copyright © 2016 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.