• J Clin Monit Comput · Aug 2002

    Closed-loop control if the inspired fraction of oxygen in mechanical ventilation.

    • Fleur Tehrani, Mark Rogers, Takkin Lo, Thomas Malinowski, Samuel Afuwape, Michael Lum, Brett Grundl, and Michael Terry.
    • Department of Electrical Engineering, California State University, Fullerton, California 92834-6870, USA. ftehrani@fullerton.edu
    • J Clin Monit Comput. 2002 Aug 1; 17 (6): 367-76.

    ObjectiveSupplemental oxygen treatment of patients on mechanical ventilation is crucial in maintaining the patients' oxygen levels in the normal range. The purpose of this study was to evaluate the effectiveness of a closed-loop controller for automatic adjustment of the fraction of inspired oxygen, FIO2. More specifically, the aim of the study was to assess the robustness of the controller in correcting hypoxemia as well as its effectiveness in prevention of hyperoxemia and oxygen toxicity.MethodsThe microprocessor-based feedback control system combines a rapid control algorithm with a proportional-integral-derivative (PID) control procedure to automatically adjust FIO2. The system is designed to prevent hypoxemia by applying a stepwise control procedure in response to rapid declines in arterial oxygen saturation while fine-tuning FIO2 and avoiding hyperoxemia by resuming to the PID control procedure when appropriate. The system includes a sophisticated safeguard unit which is designed to communicate any oxygenation problems or measurement artifacts to the medical personnel while keeping FIO2 at a safe and sufficiently high level. The control system has been tested by using computer simulations as well as animal studies.ResultsIn response to different disturbances, the arterial oxygen saturation returned to the normal safe range within less than 20 seconds, thereby avoiding any untoward effects of hypoxemia. Under steady state conditions, the variations in arterial oxygen saturation were kept within +/- 3% of the mean value. The controller corrected hypoxemia within seconds while preventing hyperoxemia, rejecting artifacts, and minimizing exposure to high concentrations of oxygen.ConclusionThe results of the study attest to the reliability of the proposed closed-loop control scheme for automatic adjustment of FIO2. Further evaluation of the controller will require testing the effectiveness of the system on different patient groups.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.