Conf Proc IEEE Eng Med Biol Soc · Jan 2010
Mechanical analysis of an oscillatory positive expiratory pressure device used in respiratory rehabilitation.
- C E Alves, L Q Nunes, and P L Melo.
- Biomedical Instrumentation Laboratory, State University of Rio de Janeiro, Brazil. caedualves@gmail.com
- Conf Proc IEEE Eng Med Biol Soc. 2010 Jan 1;2010:2477-80.
AbstractThis article aims to characterize the mechanical behavior of the Acapella Blue, a respiratory rehabilitation device designed to aid sputum clearance. In this scope, the present study initially describes in detail the peak-to-peak oscillation amplitude (App) and peak frequency (fp) behavior, as well as positive pressure level (Ppl), in the flow range more comonly found in practice. The parameters were evaluated in all 5 adjustment levels of the equipment in intervals of 50 mL/s. The device characterization has shown fp up to 23 Hz, App from 0.2 to 2.8 cmH(2)O and Ppl ranging from 1.2 to 13.5 cmH(2)O. The studied device may produce oscillation in the ranges of ciliary movements and respiratory system resonance frequency of patients with respiratory diseases. Data obtained in this work may help to optimize the use of the Acapella Blue device in respiratory rehabilitation. Suggestions for the practical use of the device are also presented.
Notes
Knowledge, pearl, summary or comment to share?