    Fifteen years experience with finger arterial pressure monitoring: assessment of the technology.

    • B P Imholz, W Wieling, G A van Montfrans, and K H Wesseling.
    • Department of Internal Medicine, Academic Medical Center, Amsterdam, Netherlands.
    • Cardiovasc. Res. 1998 Jun 1;38(3):605-16.

    AbstractWe review the Finapres technology, embodied in several TNO-prototypes and in the Ohmeda 2300 and 2300e Finapres NIBP. Finapres is an acronym for FINger Arterial PRESsure, the device delivers a continuous finger arterial pressure waveform. Many papers report on the accuracy of the device in comparison with intra-arterial or with noninvasive but intermittent blood pressure measurements. We compiled the results of 43 such papers and found systolic, diastolic and mean accuracies, in this order, ranging from -48 to 30 mmHg, from -20 to 18 mmHg, and from -13 to 25 mmHg. Weighted for the number of subjects included pooled accuracies were -0.8 (SD 11.9), -1.6 (8.3) and -1.6 (7.6) mmHg respectively. Subdividing the pooled group according to criteria such as reference blood pressure, place of application, and prototype or commercial device we found no significant differences in mean differences or SD. Measurement at the finger allows uninterrupted recordings of long duration. The transmission of the pressure pulse along the arm arteries, however, causes distortion of the pulse waveform and depression of the mean blood pressure level. These effects can be reduced by appropriate filtering, and upper arm 'return-to-flow' calibration to bring accuracy and precision within AAMI limits. For the assessment of beat-to-beat changes in blood pressure and assessment of blood pressure variability Finapres proved a reliable alternative for invasive measurements when mean and diastolic pressures are concerned. Differences in systolic pressure are larger and reach statistical significance but are not of clinical relevance. Finger arteries are affected by contraction and dilatation in relation to psychological and physical (heat, cold, blood loss, orthostasis) stress. Effects of these phenomena are reduced by the built-in Physiocal algorithm. However, full smooth muscle contraction should be avoided in the awake patient by comforting the patient, and covering the hand. Arterial state can be monitored by observing the behaviour of the Physiocal algorithm. We conclude that Finapres accuracy and precision usually suffice for reliable tracking of changes in blood pressure. Diagnostic accuracy may be achieved with future application of corrective measures.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…