Case ReportsOxygen contamination of the nitrous oxide pipeline supply.
- A Lye and R Patrick.
- Department of Anaesthesia, Royal Prince Alfred Hospital, Sydney, New South Wales.
- Anaesth Intensive Care. 1998 Apr 1;26(2):207-9.
AbstractMidogas (CIG Australia) nitrous oxide/oxygen blenders are commonly used in delivery wards in Australia. In this case report we describe an incident where a small hole in the diaphragm of the oxygen failure alarm in a Midogas nitrous oxide/oxygen blender led to retrograde flow of oxygen contaminating the nitrous oxide pipeline supply to the operating theatres and the delivery ward. This caused a reduced level of nitrous oxide to the patient in theatre, but there was no adverse outcome. However, if the oxygen pipeline pressure had been lower than the nitrous oxide pipeline pressure, the contamination would have been reversed, potentially resulting in a hypoxic mixture being delivered to many patients. Suggestions are made as to a method of prevention of this problem.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.