Comparative StudyContinuous thermodilution measurement of cardiac output: in-vitro and in-vivo evaluation.
- T Mihaljevic, L K von Segesser, M Tönz, B Leskosek, R Jenni, and M Turina.
- Clinic for Cardiovascular Surgery, University Hospital Zürich, Switzerland.
- Thorac Cardiovasc Surg. 1994 Feb 1;42(1):32-5.
AbstractThe current study was designed to evaluate a method for continuous measurement of cardiac output. The system consists of a modified pulmonary artery catheter that uses the thermodilution principle for determination of cardiac output. The evaluation was performed in vitro and in vivo. In-vitro evaluation was performed using a simple flow bench model (flow 2-9 L/min). Both continuous and bolus thermodilution methods were compared. Both methods showed good correlation with the pump flow calibrated using a volumetric tank and timer (correlation coefficient (r) for bolus thermodilution = 0.92, r for continuous thermodilution = 0.90). In-vivo evaluation was performed in six bovine experiments. Data from a total of 87 pairs of bolus versus continuous measurements were obtained. The cardiac output ranged from 1.9 to 8.9 L/min. The absolute measurement bias was not significant (mean: -0.07 L/min; 95% confidence limits: -0.87 and 0.73 L/min). The squared correlation coefficient from linear regression was 0.92. The results from this study suggest that the new continuous thermodilution measurement system for cardiac output provides accurate data in vitro and in vivo. Continuous monitoring of cardiac output adds a new dimension for evaluation of the patient's hemodynamic profile. Furthermore, significant volume load due to bolus thermodilution measurements can be avoided.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.