Comparative StudyIntraoperative accuracy of a point-of-care glucose meter compared with simultaneous central laboratory measurements.
- Boris Mraovic, Eric S Schwenk, and Richard H Epstein.
- Department of Anesthesiology, Jefferson Medical College, Thomas Jefferson University, Philadelphia, Pennsylvania 19107, USA. boris.mraovic@jefferson.edu
- J Diabetes Sci Technol. 2012 May 1;6(3):541-6.
BackgroundConcerns have been raised about the use of point-of-care (POC) glucose meters in the hospital setting. Accuracy has been questioned especially in critically ill patients. Although commonly used in intensive care units and operating rooms, POC meters were not approved by the Food and Drug Administration for such use. Data on POC glucose meter performance during anesthesia are lacking. We evaluated accuracy of a POC meter in the intraoperative setting.MethodsWe retrospectively reviewed 4,333 intraoperative records in which at least one intraoperative glucose was measured using electronic medical records at a large academic hospital. We evaluated the accuracy of a POC glucose meter (ACCU-CHEK® Inform, Roche Pharmaceuticals) based on the 176 simultaneous central laboratory (CL) blood glucose (BG) measurements that were found (i.e., documented collection times within 5 minutes). Point-of-care and central lab BG differences were analyzed by Bland-Altman and revised error grid analysis (rEGA).ResultsMean POC BG was 163.4 ± 64.7 mg/dl [minimum (min) 48 mg/dl, maximum (max) 537 mg/dl] and mean CL BG was 162.6 ± 65.1 mg/dl (min 44 mg/dl, max 502 mg/dl). Mean absolute difference between POC and CL BG was 24.3 mg/dl. Mean absolute relative difference was 16.5% with standard deviation 26.4%. Point-of-care measurements showed a bias of 0.8 relative to the corresponding CL value, with a precision of 39.0 mg/dl. Forty (23%) POC BG values fell outside the Clinical and Laboratory Standards Institute guideline and 3.4% POC measurements fell in zones C and D of the rEGA plot.ConclusionsThe tested POC glucose meter performed poorly compared to a CL analyzer intraoperatively. Perioperative clinicians should be aware of limitations of specific POC glucose meters, and routine use of POC glucose meters as sole measurement devices in the intraoperative period should be carefully considered.© 2012 Diabetes Technology Society.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.