The present study compares the cardiac output (CO) estimated by a new, non-invsive continuous Doppler device (Ultrasonic cardiac output monitor-USCOM) with that by bolus thermodilution technique (TD). Thirty post off-pump coronary artery bypass graft surgery patients were studied in this prospective nonrandomized study. Right heart CO estimation by USCOM and TD was performed and measured in quadruplet. ⋯ The mean CO was 4.63 and 4.76 Llmin as estimated by TD and USCOM respectively. For TD and USCOM, the CO had a mean bias (tendency of one technique to differ from other) of -0.13 L/min and limits of agreement (mean bias +/- 2SD) at -0.86 and 0.59 L/min. The study reveals very good agreement between the values of CO estimated by USCOM and TD.
Dheeraj Arora, Rajesh Chand, Yatin Mehta, and Naresh Trehan.
Department of Anaesthesiology and Critical Care, Escorts Heart Institute and Research Centre, New Delhi, India.
Ann Card Anaesth. 2007 Jul 1;10(2):132-6.
AbstractThe present study compares the cardiac output (CO) estimated by a new, non-invsive continuous Doppler device (Ultrasonic cardiac output monitor-USCOM) with that by bolus thermodilution technique (TD). Thirty post off-pump coronary artery bypass graft surgery patients were studied in this prospective nonrandomized study. Right heart CO estimation by USCOM and TD was performed and measured in quadruplet. A total of 120 paired observations were made. The mean CO was 4.63 and 4.76 Llmin as estimated by TD and USCOM respectively. For TD and USCOM, the CO had a mean bias (tendency of one technique to differ from other) of -0.13 L/min and limits of agreement (mean bias +/- 2SD) at -0.86 and 0.59 L/min. The study reveals very good agreement between the values of CO estimated by USCOM and TD.