    Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog.

    • Anderson F da Cunha, Sara J Ramos, Michelle Domingues, Hugues Beaufrère, Amanda Shelby, Rhett Stout, and Mark J Acierno.
    • Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA, USA.
    • Vet Anaesth Analg. 2016 Mar 1; 43 (2): 199-203.

    ObjectivesTo compare two commonly used oscillometric technologies for obtaining noninvasive blood pressure (NIBP) measurements and to determine if there is a difference in agreement between these systems and invasive blood pressure (IBP) measurements.Study DesignProspective, experimental study.AnimalsTwenty adult laboratory dogs.MethodsEach dog was anesthetized and its median caudal artery catheterized for IBP monitoring. An NIBP cuff was placed in the middle third of the antebrachium and attached to either monitor-1 or monitor-2. Four pairs of concurrent NIBP and IBP measurements were recorded with each monitor. Agreement between IBP and NIBP measurements was explored using Bland-Altman analysis, as well as the American College of Veterinary Internal Medicine (ACVIM) and Association for the Advancement of Medical Instrumentation (AAMI) guidelines for the validation of NIBP devices.ResultsBoth NIBP technologies produced results that met the ACVIM and AAMI guidelines for the validation of NIBP devices. For monitor-1, analyses of agreement showed biases of 0.2 mmHg [95% limits of agreement (LoA) -11.8 to 12.3 mmHg] in systolic arterial pressure (SAP) values, -2.6 mmHg (95% LoA -14.4 to 9.1 mmHg) in diastolic arterial pressure (DAP) values, and -2.5 mmHg (95% LoA -12.7 to 7.3 mmHg) in mean arterial pressure (MAP) values. For monitor-2, analyses of agreement showed biases of 3.4 mmHg (95% LoA -8.7 to 15.5 mmHg) in SAP values, 2.2 mmHg (95% LoA -6.6 to 10.9 mmHg) in DAP values, and 1.6 mmHg (95% LoA -5.9 to 8.9 mmHg) in MAP values.Conclusions And Clinical RelevanceMulti-function monitors can contain components from various manufacturers. Clinicians should consider whether these have been validated in the species to be monitored. Both of the technologies studied here seem appropriate for use in dogs.Published 2015. This article is a U.S. Government work and is in the public domain in the USA.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…