• Anesthesiology · Nov 1994

    Imposed work and oxygen delivery during spontaneous breathing with adult disposable manual ventilators.

    • D Hess, C Hirsch, C Marquis-D'Amico, and R M Kacmarek.
    • Respiratory Care Department Laboratory, Massachusetts General Hospital, Boston 02114.
    • Anesthesiology. 1994 Nov 1;81(5):1256-63.

    BackgroundManual ventilators (resuscitators) are used primarily to ventilate the lungs of patients lacking spontaneous ventilatory effort. However, in many settings patients are allowed to breathe through the manual ventilator. Although many aspects of manual ventilator function have been studied, very little has been reported on the use of manual ventilators during spontaneous breathing. The purpose of this study was to evaluate inspiratory and expiratory imposed work of breathing and oxygen delivery during spontaneous breathing through disposable manual ventilators.MethodsSimulated spontaneous breathing was established with a two-chambered test lung, with one chamber serving as the test chamber and the other as the driving chamber. Imposed work of breathing was evaluated with decelerating inspiratory flow at a rate of 20 breaths/min at tidal volume (VT) 0.25 1 and peak flow 40 l/min, at VT 0.5 l and peak flow 80 l/min, and VT 0.81 and peak flow 120 l/min. Flow (integrated to volume) and pressure were measured between the manual ventilator and test lung, and inspiratory and expiratory imposed work of breathing were calculated by integration of the volume-pressure curve. Oxygen concentration was measured with an oxygen analyzer placed between the manual ventilator and the test lung at 20 breaths/min, VT 0.5 l, and flow 45 l/min. An oxygen flow of 15 l/min was added to the device for all evaluations. Two of the manual ventilators had built-in positive end-expiratory pressure valves, and imposed work was evaluated at 10 cmH2O with these.ResultsThere were significant differences in imposed work between inspiration and expiration (P < 0.001) and among the three levels of ventilatory demand (P < 0.001). For each ventilatory demand, there was a significant difference in work between manual ventilator brands for inspiratory work and expiratory work (P < 0.001). At a VT of 0.5 l and peak flow of 80 l/min, the pooled inspiratory imposed work for all manual ventilators was 0.44 +/- 0.12 J/l, and the pooled expiratory imposed work was 0.29 +/- 0.05 J/l. With 10 cmH2O positive end-expiratory pressure, the inspiratory imposed work was very high (> 1 J/l). Four of the devices were unable to deliver more than 0.85 oxygen concentration at the spontaneous ventilatory pattern evaluated.ConclusionsAdult disposable manual ventilators produce a substantial imposed work of spontaneous breathing, which is increased with the addition of positive end-expiratory pressure. With some manual ventilators, a high oxygen concentration may not be delivered during spontaneous breathing. We recommend that patients not be allowed to spontaneously breathe through disposable manual ventilators.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.