• Anesthesia and analgesia · Feb 1991

    Comparative Study

    Saline-filled cuffs help prevent laser-induced polyvinylchloride endotracheal tube fires.

    • M B Sosis and F X Dillon.
    • Department of Anesthesiology, Rush-Presbyterian-St. Luke's Medical Center, Chicago, Illinois 60612.
    • Anesth. Analg. 1991 Feb 1;72(2):187-9.

    AbstractTo determine whether the filling of tracheal tube cuffs with saline would decrease their combustibility during laser surgery, 20 polyvinylchloride tracheal tubes were studied. The cuffed end of each tracheal tube was inserted into the neck of an empty flask, and the tube and flask were flushed with oxygen for 5 min before cuff inflation. Ten tracheal tubes had their cuffs inflated with air, and 10 were inflated with saline. A Lasersonics LS880 CO2 laser, set to 5 W for five of each of the two types of filled cuffs and to 40 W for the other pair of five tubes, was fired continuously at the cuffs for up to 1 min. No combustion occurred at the 5-W setting. The times to cuff perforation when the laser was set at 5 W were (mean +/- SD) 1.00 +/- 0.83 and 4.21 +/- 3.91 s for the air- and saline-filled cuffs, respectively, a difference that was not statistically significant. The time to deflation of the saline-filled cuff (104.6 +/- 67.5 s) was, however, significantly longer than that of the air-filled cuff (2.59 +/- 1.97 s). When the tracheal tube cuffs were exposed to 40-W laser radiation, the cuff and adjacent tube shaft ignited in all cases when the cuffs were inflated with air, but only in one of five cases when the cuffs were filled with saline (P less than 0.05). The filling of tracheal tube cuffs with saline provides simple, moderately effective partial protection of the cuff of endotracheal tubes during CO2 laser airway surgery.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…