Comparative Study
Fluid flow through dural puncture sites. An in vitro comparison of needle point types.
- R H Cruickshank and J M Hopkinson.
- Department of Anaesthetics, Royal Hallamshire Hospital, Sheffield.
- Anaesthesia. 1989 May 1;44(5):415-8.
AbstractLeakage of artificial cerebrospinal fluid through human dura was measured in vitro after puncture by spinal needles. Fluid loss tailed off in all cases and ceased within 5 minutes in 10% of punctures made with a 22-gauge needle, 28% made with a 26-gauge and 65% made with a 29-gauge needle (p less than 0.05). The fluid loss was not reduced by alignment of the level of the needle parallel to the longitudinal direction of the fibres. Whitacre point needles of 22 gauge produced lower fluid loss than a 22-gauge Quincke point needle inserted across the fibres (p less than 0.05). Leakage rate was related to needle size, but not related to the alignment of a Quincke point. Little or no leakage occurred with 29-gauge needles.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.