• Eur J Anaesthesiol · Jan 1991

    Hemocue--a new haemoglobinometer in the clinic.

    • W K Be, H E Kerkkamp, and L H Booij.
    • Department of Anesthesiology, St. Radboudhospital, University of Nijmegen, The Netherlands.
    • Eur J Anaesthesiol. 1991 Jan 1;8(1):55-8.

    AbstractThe accuracy of the Hemocue, a portable haemoglobinometer, was tested on 29 patients undergoing open-heart surgery, and the results were compared with the haemoglobin values obtained from the hospital laboratory. Special attention was paid to the lower range of haemoglobin values and the simplicity and speed of operating the meter. The correlation between the methods was good (r = 0.965). The limits of agreement were 0.0323 +/- 0.705 mmol l-1 (95% confidence level).

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.