- Dietrich Gravenstein, Harshdeep Wilkhu, Edwin B Liem, Stuart Tilman, and Samsun Lampotang.
- Department of Anesthesiology, University of Florida College of Medicine, Gainesville, Florida, USA. dgravenstein@anest.ufl.edu
- Anesth. Analg. 2007 Apr 1;104(4):860-2.
AbstractWe describe three cases of previously unreported failures of the Bag-Ventilator Switch in Aestiva/5 anesthesia machines (GE Healthcare/Datex-Ohmeda, Madison, WI). Each failure mode produced a large breathing-circuit leak. Examination of the switches revealed a cracked toggle actuator, residue build-up, and a cracked selector switch housing as causes for the failures. When a leak with no visible cause develops, consider advancing the mode selector switch fully to its mechanical limit or consider that the toggle actuator or its anchoring mechanism may have failed. These cases demonstrate that it is imperative to always be prepared to immediately use an alternate method for ventilation. Cases describing failure to ventilate due to sudden equipment malfunction underscore the need to always have functioning backup ventilation equipment available.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.