- R von Roemeling, R M Lanning, and F A Eames.
- Medical Oncology Division, Albany Medical College, Union University, Albany, NY.
- J Magn Reson Imaging. 1991 Jan 1;1(1):77-81.
AbstractThe influence of magnetic resonance (MR) imaging on the integrity and function of an implantable, programmable drug infusion pump and the distortion of the image by this device were tested. Six identical pumps were tested in magnetic fields of 1.5 T. Pump memory was not affected by the procedure. The pump rotor stalled (no infusion) as long as the device was within the magnetic field. Drug delivery resumed as programmed after the pump was removed from the field. No structural damage to the electrical or mechanical pump components was detected. The pump caused a circular image artifact (a signal dropout) within 8-10 cm of the device. The authors conclude that MR imaging is accurate if the area of interest is at least 10 cm from the pump, and that with awareness of temporary infusion cessation, it is safe to perform MR imaging in patients with this implanted pump.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.