-
- P D Slinger, W A Scott, and A P Kliffer.
- Department of Anaesthesia, Montreal General Hospital, Quebec, Canada.
- Can J Anaesth. 1990 Mar 1;37(2):258-61.
AbstractA case of intraoperative awareness during a thoracotomy is described. The patient's recall coincided with an intraoperative period during which a Siemens 900B ventilator and a Siemens 952 isoflurane vaporiser were used. Subsequent assessment of this equipment with an anaesthetic agent analyzer revealed that, at the ventilator settings which had been used, the delivered anaesthetic vapour concentration varied greatly from the vaporizer settings. This problem eventually was traced to a malfunctioning inlet control valve on the ventilator. This complication may have been prevented if the end-tidal anaesthetic concentration had been monitored intraoperatively.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.