Comparative Study
Comparison of four different emergency airway access equipment sets on a human patient simulator.
- B S Vadodaria, S D Gandhi, and A K McIndoe.
- Anaesthetist, Northern General Hospital, Herries Road, Sheffield S5 7AU, UK. bvadodaria@aol.com
- Anaesthesia. 2004 Jan 1;59(1):73-9.
AbstractThis study used a patient simulator to study the ease of use and efficacy of four currently available cricothyroidotomy sets. We assessed the success of insertion of each piece of equipment and measured the subsequent adequacy of oxygenation and ventilation. We also examined the complications encountered using each set. We found that there was a 100% success rate of achieving an adequate airway within acceptable time limits using the 'Quiktrach' and 'Melker' sets, with good airway patency and ease of ventilation. There was an unacceptably high failure rate in achieving a patent airway when using the 'Transtracheal airway catheter with ENK-flow modulator' and 'Patil's airway'. The pre-assembled and user-friendly 'Quiktrach' set provided the fastest and most effective means of oxygenation in the simulated patient requiring an emergency surgical airway.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.