• Critical care medicine · May 1996

    Accuracy and reliability of disposable pressure transducers coupled with modern pressure monitors.

    • R M Gardner.
    • Department of Medical Informatics, LDS Hospital, Salt Lake City, UT 84143, USA.
    • Crit. Care Med. 1996 May 1;24(5):879-82.

    ObjectiveTo determine the bedside accuracy of direct patient pressure monitoring when used with new and clinically used disposable blood pressure (BP) transducers.DesignProspective study.SettingLaboratory bench and critical care units in an adult and children's hospital.SubjectsSeventy-five bedside patient monitors (25 Marquette Electronics, 25 Spacelab Medical, and 25 Hewlett-Packard), and 100 disposable transducers (50 from Utah Medical Products and 50 from Abbott Critical Care Systems [25 new, 25 clinically used of each manufacturer]) were tested.InterventionsNone.Measurements And Main ResultsA +/- 2% accuracy requirement for bedside monitors and the +/- 3% American National Standards Institute accuracy standard for disposable BP transducers were used. To test the accuracy of the bedside monitors, a certified transducer simulator was used to apply 100 mm Hg to each bedside monitor. To test the accuracy of the disposable BP transducers, a very accurate (+/- 0.05%) pneumatic dead weight tester was used to apply pressures to the transducer. A digital power supply and a 6 1/2 digit voltmeter were used. The average output of the bedside monitors when 100 mm Hg was applied was 99.90 +/- 0.83 mm Hg, with the worst cases being 98 and 103 mm Hg. For all 100 disposable pressure transducers, the average output was 100.03 +/- 0.55 mm Hg, with the worst cases being 98.53 and 101.36 when 100 mm Hg was applied. There was no important difference in the accuracy of the transducers obtained from the two vendors nor whether the transducers had been used clinically.ConclusionsAll disposable BP transducers tested were much more accurate than the American National Standards Institute standard for accuracy. Even the worst case transducers were twice as accurate as required by the American National Standards Institute standard. Only one bedside monitor was outside the +/- 2% accuracy range (103 mm Hg). Based on these findings, this author recommends that fixed calibration disposable transducers and fixed calibration bedside pressure monitoring systems be used. The clinical risks of air embolism and infection from the calibrating mercury manometer and the complexity of the calibration task are the overriding factors for making these recommendations.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.