Investigative radiology · May 2004
Neurostimulation system used for deep brain stimulation (DBS): MR safety issues and implications of failing to follow safety recommendations.
- Ali R Rezai, Michael Phillips, Kenneth B Baker, Ashwini D Sharan, John Nyenhuis, Jean Tkach, Jaimie Henderson, and Frank G Shellock.
- The Cleveland Clinic Foundation, Cleveland, OH, USA.
- Invest Radiol. 2004 May 1;39(5):300-3.
AbstractThe use of magnetic resonance imaging (MRI) in patients with neurostimulation systems used for deep brain stimulation requires the utmost care, and no individual should undergo an MR examination in the absence of empirical evidence that the procedure can be performed safely. The risks of performing MRI in patients with neurostimulators include those associated with heating, magnetic field interactions, induced currents, and the functional disruption of these devices. The exact safety recommendations for the particular neurostimulation system with regard to the pulse generator, leads, electrodes, operational conditions for the device, the positioning of these components, and the MR system conditions must be carefully followed for MRI. As highlighted by 2 recent accidents, the failure to strictly follow safety recommendations (eg, use a 1.5-T MR system with a send/receive head radiofrequency coil only; limit the specific absorption rate to 0.4 W/kg; etc.) may result in serious, temporary, or permanent injury to the patient including the possibility of transient dystonia, paralysis, coma, or even death.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.