• J Clin Monit Comput · May 1999

    Amount of air infused to patient increases as fluid flow rates decrease when using the Hotline HL-90 fluid warmer.

    • S Woon and P Talke.
    • University of California, Los Angeles 94143-0648, USA.
    • J Clin Monit Comput. 1999 May 1; 15 (3-4): 149-52.

    ObjectiveThe intraoperative use of fluid warming devices has been recommended to avoid perioperative hypothermia and related adverse outcomes. To evaluate whether these devices might introduce risks of their own, we measured the volume of air escaping from a warmed intravenous solution that might be delivered to a patient.MethodsIn an operating room maintained at 19-19.5 degrees C, we tested an HL-90 Hotline fluid warmer with the L-70 fluid-warming set. One liter of lactated Ringer's solution was infused at flow rates of 150, 300, 500 and 3400 ml/h. The air that formed within the L-70 tubing during infusion was collected in a bubble trap placed at the end of the L-70 tubing. The volume of air in the bubble trap was measured. Twelve separate measurements were obtained at each flow rate. One additional study (n = 8) was performed using the L-10 Gas Vent to determine whether this equipment might reduce the volume of air infused when fluid flow rate was 300 mL/h. The volume of air collected at each flow rate was compared using ANOVA.ResultsVolume of air increased significantly from 1.0 +/- 0.2 mL to 2.9 +/- 0.4 ml as flow rate decreased from 3400 ml/h to 150 ml/h (p < 0.0001). The L-10 gas eliminator was ineffective in reducing the amount of air infused.ConclusionsWe conclude that the use of the Hotline fluid warmer can result in infusion of air into the patient, introducing possible risk of air embolism.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…