• Journal of anesthesia · Jun 2014

    Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

    • Daisuke Toyoda, Rie Yasumura, Mitsue Fukuda, Ryoichi Ochiai, and Yoshifumi Kotake.
    • Department of Anesthesiology, Toho University Medical Center Ohashi Hospital, 2-17-6 Ohashi, Meguro, Tokyo, 153-8515, Japan.
    • J Anesth. 2014 Jun 1;28(3):463-6.

    AbstractThe purpose of this prospective study was to evaluate the accuracy and trending ability of a four-wavelength pulse-total hemoglobinometer that continuously and noninvasively measures hemoglobin in surgical patients. With IRB approval and informed consent, spectrophotometric hemoglobin (SpHb) was measured with a pulse-total hemoglobinometer manufactured by Nihon Kohden Corp (Tokyo, Japan) and compared to the CO-oximeter equipped with blood gas analyzer. Two hundred twenty-five samples from 56 subjects underwent analysis. Bland-Altman analysis revealed that the bias ± precision of the current technology was 0.0 ± 1.4 g/dl and -0.2 ± 1.3 g/dl for total samples and samples with 8 < Hb < 11 g/dl, respectively. The percentages of samples with intermediate risk of therapeutic error in error grid analysis and the concordance rate of 4-quadrant trending assay was 17 % and 77 %, respectively. The Cohen kappa statistic for Hb < 10 g/dl was 0.38, suggesting that the agreement between SpHb and CO-oximeter-derived Hb was fair. Collectively, wide limits of agreement, especially at the critical level of hemoglobin, and less than moderate agreement against CO-oximeter-derived hemoglobin preclude the use of the pulse-total hemoglobinometer as a decision-making tool for transfusion.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.