• Critical care medicine · Sep 1988

    Comparative Study

    Some infant ventilators do not limit peak inspiratory pressure reliably during active expiration.

    • H Kirpalani, R Santos-Lyn, and R Roberts.
    • Divisions of Neonatology, Hospital for Sick Children, Toronto, Ontario, Canada.
    • Crit. Care Med. 1988 Sep 1;16(9):880-3.

    AbstractIn order to minimize barotrauma in newborn infants with respiratory failure, peak inspiratory pressures should not exceed those required for adequate gas exchange. We examined whether four commonly used pressure-limited, constant flow ventilators limit pressure reliably during simulated active expiration against the inspiratory stroke of the ventilator. Three machines of each type were tested at 13 different expiratory flow rates (2 to 14 L/min). Flow-dependent pressure overshoot above a dialed pressure limit of 20 cm H2O was observed in all machines. However, the magnitude differed significantly between ventilators from different manufacturers (p = .0009). Pressure overshoot above 20 cm H2O was consistently lowest in the Healthdyne (0.8 cm H2O at 2 L/min, 3.6 cm H2O at 14 L/min) and highest in the Bourns BP200 (3.0 cm H2O at 2 L/min, 15.4 cm H2O at 14 L/min). We conclude that peak inspiratory pressure overshoots on pressure-limited ventilators occur during asynchronous expiration. This shortcoming may contribute to barotrauma in newborn infants who "fight" positive-pressure ventilation.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.