• Acta Anaesthesiol Scand · Jan 1992

    A new ventilator converter with the Siemens Servo Ventilator--evaluation in a lung model.

    • J F Hesselvik, M Bengtsson, and A Johnson.
    • Department of Anaesthesiology, University Hospital, Linköping, Sweden.
    • Acta Anaesthesiol Scand. 1992 Jan 1;36(1):75-9.

    AbstractA ventilator converter device (Anmedic) for connecting a non-rebreathing ventilator (Servo Ventilator 900 B; Siemens-Elema) to a circle anaesthesia system was evaluated in a lung model. Recorded minute ventilation was slightly lower than dialled in most cases. We furthermore found inadequate expiratory expansion of the ventilator converter bellows, with progressive loss of tidal volume and consequently recorded minute volume, when fresh gas flow to the circle system was low (1 l.min-1), expiratory time was short (less than 40%) and respiratory obstruction was present.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.