• American heart journal · Dec 1996

    Active compression-decompression versus standard cardiopulmonary resuscitation in a porcine model: no improvement in outcome.

    • K B Kern, G Figge, R W Hilwig, A B Sanders, R A Berg, C W Otto, and G A Ewy.
    • Department of Medicine, University of Arizona College of Medicine, Tucson 85724, USA.
    • Am. Heart J. 1996 Dec 1;132(6):1156-62.

    AbstractActive compression-decompression cardiopulmonary resuscitation (CPR) is a new innovative basic life-support technique during which the anterior chest wall is actively decompressed by a suction device. CPR techniques were studied in 36 swine to test the hypothesis that active compression-decompression CPR improves coronary perfusion pressure, myocardial blood flow during CPR, and 24-hour survival. After 30 seconds of untreated ventricular fibrillation, CPR was begun and continued for 12.5 minutes by one of the three following methods: (1) active compression-decompression CPR with a suction device modified to include a precision force transducer; (2) standard CPR performed with a force transducer device; and (3) standard manual CPR performed without a force transducer device. CPR-generated coronary perfusion pressure, myocardial blood flow, and the force of compression were measured at 3 and 10 minutes of resuscitation effort. Initial return of spontaneous circulation, 24-hour survival, and trauma scores were also evaluated. Active compression-decompression CPR produced consistently better results than did standard CPR performed with a force transducer, but not better than standard CPR performed manually without a force transducer. The use of a force-measuring device with standard CPR may compromise hemodynamic response and outcome.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…