Anesthesia and analgesia · Jan 2012
Case ReportsCase report: automated machine checkout leaves an internal gas leak undetected: the need for complete checkout procedures.
- Timothy S Eng and Marcel E Durieux.
- Department of Anesthesiology, University of Virginia, Charlottesville, 22908-0710, USA.
- Anesth. Analg.. 2012 Jan 1;114(1):144-6.
AbstractWe report a complete internal fresh gas flow disconnect within a Dräger Fabius GS anesthesia machine without any alarms being triggered. This was undetected primarily because of an incomplete machine checkout in which the step of ensuring proper gas flows by using a "test lung" was omitted. Machine-specific factors, however, also contributed to prevent diagnosis: (1) the machine passed its leak test because the flowmeter bobbin (i.e., floating ball) sealed the flowmeter when back pressure was applied; (2) the mechanical ventilator entrains room air, thus functioning in the absence of fresh gas flow; and (3) the electronic flow sensors functioned "appropriately" because the leak was downstream. Despite the advent of highly automated machines, manual checkout procedures remain crucial to minimizing undiagnosed failures.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.