• Artificial organs · May 2013

    Comparative Study

    New generation extracorporeal membrane oxygenation with MedTech Mag-Lev, a single-use, magnetically levitated, centrifugal blood pump: preclinical evaluation in calves.

    • Tatsuki Fujiwara, Eiki Nagaoka, Taiju Watanabe, Naoto Miyagi, Takashi Kitao, Daisuke Sakota, Taichi Mamiya, Tadahiko Shinshi, Hirokuni Arai, and Setsuo Takatani.
    • Department of Cardiovascular Surgery, Tokyo Medical and Dental University, Bunkyo-ku, Tokyo, Japan.
    • Artif Organs. 2013 May 1;37(5):447-56.

    AbstractWe have evaluated the feasibility of a newly developed single-use, magnetically levitated centrifugal blood pump, MedTech Mag-Lev, in a 3-week extracorporeal membrane oxygenation (ECMO) study in calves against a Medtronic Bio-Pump BPX-80. A heparin- and silicone-coated polypropylene membrane oxygenator MERA NHP Excelung NSH-R was employed as an oxygenator. Six healthy male Holstein calves with body weights of about 100 kg were divided into two groups, four in the MedTech group and two in the Bio-Pump group. Under general anesthesia, the blood pump and oxygenator were inserted extracorporeally between the main pulmonary artery and the descending aorta via a fifth left thoracotomy. Postoperatively, both the pump and oxygen flow rates were controlled at 3 L/min. Heparin was continuously infused to maintain the activated clotting time at 200-240 s. All the MedTech ECMO calves completed the study duration. However, the Bio-Pump ECMO calves were terminated on postoperative days 7 and 10 because of severe hemolysis and thrombus formation. At the start of the MedTech ECMO, the pressure drop across the oxygenator was about 25 mm Hg with the pump operated at 2800 rpm and delivering 3 L/min flow. The PO2 of the oxygenator outlet was higher than 400 mm Hg with the PCO2 below 45 mm Hg. Hemolysis and thrombus were not seen in the MedTech ECMO circuits (plasma-free hemoglobin [PFH] < 5 mg/dL), while severe hemolysis (PFH > 20 mg/dL) and large thrombus were observed in the Bio-Pump ECMO circuits. Plasma leakage from the oxygenator did not occur in any ECMO circuits. Three-week cardiopulmonary support was performed successfully with the MedTech ECMO without circuit exchanges. The MedTech Mag-Lev could help extend the durability of ECMO circuits by the improved biocompatible performances.© 2013, Copyright the Authors. Artificial Organs © 2013, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.