• Chinese Med J Peking · Aug 2011

    Accuracy of the ultrasonic cardiac output monitor in healthy term neonates during postnatal circulatory adaptation.

    • Shao-ru He, Cheng Zhang, Yu-mei Liu, Yun-xia Sun, Jian Zhuang, Ji-mei Chen, Veronica M Madigan, Brendan E Smith, and Xin Sun.
    • Department of Neonatology, Guangdong Academy of Medical Sciences/Guangdong General Hospital, Guangzhou, Guangdong 510080, China.
    • Chinese Med J Peking. 2011 Aug 1;124(15):2284-9.

    BackgroundEchocardiography is regarded as a gold standard for measuring hemodynamic values. The ultrasonic cardiac output monitor (USCOM) is a new method for measuring hemodynamics and could provide non-invasive point of care guidance. So far, there are no published USCOM reference values for neonates, nor has USCOM's accuracy been established in this population. We aimed to determine the accuracy and clinical utility of the USCOM in healthy neonates relative to published echocardiographic data, to establish normal hemodynamic parameters that it measures, and to assess the possible role of USCOM as an alternative to echocardiography as a trend monitor.MethodsRight and left heart hemodynamics of 90 normal neonates were measured during circulatory adaptation over the first three days of life using the USCOM and automated oscillotonometry.ResultsHeart rate showed a significant decline from days one to three, from 126 to 120 (P < 0.001). Systolic, diastolic and mean arterial pressures all increased significantly from 66 to 71 mmHg, 33 to 38 mmHg and 44 to 49 mmHg, respectively (P < 0.001 in each case). Right ventricular cardiac index (RV-CI) showed no change with a mean of 5.07 L × min(-1) × m(-2). Left ventricular cardiac index (LV-CI) declined from 3.43 to 3.00 L × min(-1) × m(-2) (P < 0.001). RV-CI exceeded LV-CI on all three days by a mean of 61%. The systemic vascular resistance index (SVRI), based on LV-CI, increased significantly over the three days from 1083 to 1403 dyne × sec × cm(-5) × m(2) (P < 0.001).ConclusionsNormal neonatal hemodynamic values, as indicated by USCOM, were established. LV-CI measurement showed excellent agreement with published echocardiographic studies. RV-CI was constant and exceeded LV-CI for all three days of this study. It may be falsely high due to flow velocity measurement errors arising from the pulmonary branch arteries, and may represent a limitation of the USCOM method. The progressive rise of arterial pressure and SVRI despite a declining LV-CI may indicate functional closure of the ductus arteriosus, with the greatest change occurring within the first 24 hours. Evidence of closure of the foramen ovale was not observed.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…