    Four methods of measuring tidal volume during high-frequency oscillatory ventilation.

    • David N Hager, Mathew Fuld, David W Kaczka, Henry E Fessler, Roy G Brower, and Brett A Simon.
    • Division of Pulmonary and Critical Care Medicine, Department of Anesthesiology and Critical Care Medicine, Johns Hopkins University, Baltimore, MD, USA.
    • Crit. Care Med. 2006 Mar 1;34(3):751-7.

    ObjectiveAssess the accuracy of four different methods of measuring tidal volume during simulated high-frequency oscillatory ventilation.DesignIn vitro study.SettingResearch laboratory.SubjectsThree differential pressure pneumotachometers, a modified Pitot tube, an ultrasound flowmeter, and an adult hot wire anemometer.InterventionsEach device was placed in series with a Sensormedics 3100B high-frequency ventilator and an 8.0-mm endotracheal tube attached to a 48.9-L plethysmograph. Inspiratory/expiratory ratio was fixed at 1:1 and mean airway pressure at 10 cm H2O. Tidal volumes were calculated at each combination of frequency (f: 3, 4, 6, 8, 10, 12 Hz) and pressure amplitude (DeltaP: 30, 60, 90 cm H2O) by digital integration of the sampled flow signals from each device and compared with those calculated from pressure changes within the plethysmograph. The protocol was repeated after incorporation of frequency-specific calibrations to the flow-measuring algorithm of each device. The hot wire anemometer was further evaluated at Fio2 of 1.0, 37 degrees C, 80% humidity, mean airway pressure of 20 cm H2O, and an inspiratory/expiratory ratio of 1:2.Measurements And Main ResultsTidal volumes were 36-305 mL. Bland-Altman analysis demonstrated that each device exhibited systematic bias before frequency-specific adjustment. After frequency-specific adjustment of the flow-measuring algorithm, the two most accurate and precise devices were the Hans Rudolph pneumotachometer, which exhibited a mean error of 0.2% (95% confidence interval, -3.0% to 3.4%), and the hot wire anemometer, which had a mean error of -1.1% (95% confidence interval, -5.5% to 3.3%). The hot wire anemometer remained accurate at Fio2 1.0, 37 degrees C, 80% humidity, mean airway pressure of 20 cm H2O, and an inspiratory/expiratory ratio of 1:2.ConclusionsTidal volume can be measured during high-frequency oscillatory ventilation using a variety of techniques. Frequency-specific calibration improves the accuracy and precision of tidal volume measurements. Hot wire anemometry exhibits stable performance characteristics across the range of temperature, humidity, Fio2, and inspiratory/expiratory ratios encountered clinically, has a small deadspace, is unaffected by mean airway pressure, and is therefore suitable for clinical applications.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.