• J Orthop Trauma · Oct 2012

    The position and number of screws influence screw perforation of the humeral head in modern locking plates: a cadaver study.

    • Johannes B Erhardt, Karl Stoffel, Jorg Kampshoff, Nicole Badur, Piers Yates, and Markus S Kuster.
    • Department of Orthopaedic Surgery, Cantonal Hospital St. Gallen, St. Gallen, Switzerland. johannes.erhardt@kssg.ch
    • J Orthop Trauma. 2012 Oct 1;26(10):e188-92.

    ObjectivesScrew perforation of the humeral head in locking plate osteosynthesis occurs in up to 30% of cases. The current study compared different fixation possibilities (eg, number and position of screws) to reduce screw perforation in the humeral head.MethodsA humeral head fracture with a missing medial support was created in 30 fresh-frozen cadavers and fixed with a polyaxial locking plate (NCB PH; Zimmer, Warsaw, IN). The constructs were loaded with increasing force and the number of cycles until screw perforation was recorded. Four different fixation methods were tested: group 1 five screws with fixed angle, group 2 five screws in polyaxial position according to bone strength, group 3 three screws, and group 4 five screws with 1 as an inferomedial support screw.ResultsMore screws in the humeral head significantly increased the number of cycles before screw perforation. An inferomedial support screw further increased the number of cycles. Polyaxial screw placement compared with fixed-angle placement had no effect on the screw perforation phenomenon.ConclusionsWe recommend to position an inferomedial support screw, and at least 5 screws in the head fragment, when using a locking plate in proximal humerus fractures with disrupted medial hinge.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…