Comparative StudyEvaluation of ventilators used during transport of ICU patients -- a bench study.
- G Zanetta, D Robert, and C Guérin.
- Service de Réanimation Médicale et Assistance Respiratoire, Hôpital de la Croix Rousse, 103 grande rue de la Croix-Rousse, 69004 Lyon, France.
- Intensive Care Med. 2002 Apr 1;28(4):443-51.
ObjectivesTo evaluate portable ventilators.Design And SettingsBench study.Materials And MethodsFive portable ventilators used for transporting ICU patients [Osiris 1, (ventilator a), Osiris 2, (ventilator b), Oxylog 1000, (ventilator c), Oxylog 2000, (ventilator d), AXR1a, (ventilator e)] and three ICU ventilators which can be used for this purpose [Horus, (ventilator f), T-Bird, (ventilator g), and SV 300, (ventilator h)] were compared using a test lung regarding: 1) their capability to maintain set tidal volumes (V(T)) of 300 ml, 500 ml, and 800 ml under a normal condition A [resistance (R) 5 cmH(2)O/l/s and compliance (C) 100 ml/cmH(2)0] and two abnormal conditions B (R 20-C 30) and C (R 50-C 100); 2) trapped volume (expired V(T)relative to inspired V(T)at 0.7 s, 1 s, and 1.4 s), an estimate of the expiratory resistance of both circuit and valve; and 3) the triggering system assessed from the measurements of Delta t, Delta P for two inspiratory efforts at a PEEP of 0 cmH(2)0 and 5 cmH(2)0 in ventilators b, d, f, g, and h. Flow and airway pressure were measured with an independent physiologic recording system.Results1) V(T). For ventilators a-h, the mean+/-SD changes of a set V(T)of 300 ml were -2.6+/-0.2%, -9.7+/-0.2%, 0+/-0%, -6.1+/-0.2%, 1.0+/-0.3%, -2.1+/-1.7%, 0.3+/-0%, and -1.3+/-0.1% ( P<0.001), respectively, during condition B relative to A. Similar results were obtained for a V(T)of 500 ml and 800 ml and during condition C relative to A; 2) Trapped volume. For ventilators a-h, trapped volume averaged 1+/-1%, 20+/-0%, 30+/-0.4%, 20+/-1%, 1+/-0%, 19+/-0%, 15+/-0%, and 14+/-0% at 0.7 s ( P <0.001) and 0.6+/-0%, 5+/-0%, 0.5+/-0%, 0+/-0%%, 0+/-0%, 0.6+/-0%, 0+/-0%, and 0+/-0% at 1.4 s ( P=NS); and 3) the triggering system of Oxylog 2000 was poor whereas it was of good quality for Horus, T-Bird, SV 300, and Osiris 2.ConclusionsThe small portable ventilators presently investigated varied between each other and were less accurate than ICU ventilators.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.