• Europace · Mar 2007

    Case Reports

    Integrated home monitoring predicts lead failure in a pacemaker dependent 4-year-old girl.

    • P A Zartner, R P Handke, A M Brecher, and M B E Schneider.
    • German Paediatric Heart Centre Sankt Augustin, Department for Cardiology, Arnold-Janssen-Str. 29, D-53757 Sankt Augustin, Germany. p.zartner@asklepios.com
    • Europace. 2007 Mar 1;9(3):192-3.

    AbstractA 4-year-old girl with post-surgical complete atrioventricular block received an epicardial dual chamber pacemaker system. During further growth intermittent exit block occurred, first misinterpreted as neurological seizures. The epicardial lead was replaced using a transvenous approach, and a pacemaker with an integrated home monitoring facility was implanted. After her discharge, a rise in the pacing threshold automatically initiated an event message. On the basis of this information, the patient was called in and imminent dislodgement of the ventricular lead was diagnosed by x-ray. The lead was repositioned and was found stable over 1-year follow-up.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…