• Eur. Respir. J. · Nov 1996

    Comparative Study

    A laboratory comparison of four positive pressure ventilators used in the home.

    • I E Smith and J M Shneerson.
    • The Respiratory Support and Sleep Centre, Papworth Hospital, Cambridge, UK.
    • Eur. Respir. J. 1996 Nov 1;9(11):2410-5.

    AbstractFour brands of ventilator used for noninvasive positive pressure ventilation in the home were assessed to determine their performance on a patient simulator. We tested the tidal volume (VT) preset Companion 2801 (Puritan Bennett), minute volume preset Monnal D (Taema), and two pressure preset ventilators, the Nippy (Friday Medical) and the Bilevel Positive Airway Pressure (BiPAP) (Respironics). A patient simulator was employed to investigate the relationships between VT, peak airway pressure (PAP) and mean airway pressure (MAP), the responses to an additional leak in the circuit and patient effort of a variable duration, which was modelled using a negative pressure pump. For equivalent VTs, >300 mL, the Monnal D and Nippy generated a PAP up to 100% greater than the Companion 2801 and BiPAP. When an additional leak was introduced to the circuit, the VT of the Companion 2801 and Monnal D fell by >50%, while the Nippy and the BiPAP responded by increasing flow and maintained VT close to the level achieved with no leak. When the ventilators were triggered by a simulated patient effort of 0.25 s duration with a frequency 33% greater than that of the ventilator, the minute volume increased by 41% for the Companion 2801, by 18% for the Monnal D (no change expected), and by 17% for the Nippy (less than expected), and fell by 7% for the BiPAP due to irregular triggering. When patient effort was prolonged to 1.5 s, a similar length to the inspiratory time of the ventilators, there was no further change in the minute volume of the Companion 2801 and Monnal D, while that of the Nippy and of the BiPAP increased by 38 and 71%, respectively, compared to baseline. These results show that distinct brands of ventilator respond to changes in the patient and patient circuit in different ways, which are not always predictable from a simple description of their operating principles. This should be borne in mind when choosing a positive pressure ventilator for noninvasive ventilation.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…