- W M Coplin, G E O'Keefe, M S Grady, G A Grant, K S March, H R Winn, and A M Lam.
- Department of Neurological Surgery, Harborview Medical Center, University of Washington, Seattle 98104, USA.
- Neurosurgery. 1998 Mar 1;42(3):533-9; discussion 539-40.
ObjectiveTo address the accuracy of a bedside jugular bulb oxygen saturation (SjO2) catheter monitor (Baxter-Edwards, Santa Ana, CA) versus in vitro co-oximetry measurements in the intensive care unit (ICU).MethodsBy prospective protocol, we compared blood gas measurements with simultaneously recorded continuous bedside oximetric monitor values for 31 ICU patients with traumatic brain injury undergoing jugular bulb catheter monitoring. For suboptimal fiberoptic light signal quality indices, the catheter was repositioned, flushed, or both before drawing the sample for in vitro measurement. Laboratory and bedside monitor data were examined for association using the chi2 and paired t tests and a linear regression model.ResultsWe assessed 195 samples (median, 5 per patient; range, 1-14) who were monitored an average of 3.4 (range, 1-6) days. The in vivo monitor (range, 32-94%) and in vitro co-oximetry (range, 38-93%) values had acceptable correlation (y = 0.94x + 4.4, r2 = 0.80). For bedside monitor detection of jugular bulb desaturation (SjO2 < 50% for 10 min), the kappa statistic was 0.35, the sensitivity was 45 to 50%, and the specificity was 98 to 100%.ConclusionContinuous ICU SjO2 monitoring correlates significantly with in vitro values, but less so than previously described during intracranial surgery. Although sensitivity of the bedside monitor to detect confirmed desaturations remains an issue, the high specificity indicates that it is less of a concern that patients may be misdiagnosed as having desaturations resulting in unnecessary interventions. Nonetheless, suspected jugular bulb desaturation should be verified before taking therapeutic actions.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.