Comparative Study
Estimation of desflurane concentration using isoflurane channel in optical infrared analyzer.
- B C Tsui, D T Jolly, and A S Clanachan.
- Department of Anesthesia, University of Alberta Hospitals, Edmonton, Canada.
- Can J Anaesth. 1999 Feb 1;46(2):192-5.
PurposeTo estimate desflurane concentration on the isoflurane channel in an optical infrared analyzer using a simple regression equation.MethodsDesflurane in concentrations of 0% to 3% in 0.5% increments and 3% to 12% in 1% increments in 2 L.min-1 oxygen was delivered simultaneously to an Ohmeda 5250 RGM desflurane channel, an Ohmeda 5250 RGM isoflurane channel, and a Datex Capnomac Ultima isoflurane channel at room temperature and atmospheric pressure. For each concentration increment, the displayed gas concentrations were recorded. By comparing the readings from the desflurane channel of Ohmeda RGM and the isoflurane channels from Ohmeda RGM and Datex Capnomac Ultima respectively, the linear regression relationship and the slope of the fitted line (conversion factor) between two channels were obtained. Similar measurements were performed using 2 L.min-1 mixture of nitrous oxide 50% and oxygen 50%. The measurements were repeated with different monitors three months later.ResultsAll four analysers tested were linear (r2 > 0.9) for measuring desflurane using isoflurane channels over the range of concentrations studied on two different days. The accuracy of the estimation using the mean conversion factor of the four monitors was within 10% error from the readings of the commercially available desflurane channel analyzer. There was no noticeable effect on the slope (conversion factor) of the linear regression with O2 100% or 50/50 mixture of N2O and O2.ConclusionThe concentration of desflurane can be estimated by a simple conversion factor using an isoflurane channel of an infrared system.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.