    Accuracy of methemoglobin measurements: comparison of six different commercial devices and one manual method.

    • J Dötsch, S Demirakça, D Hahn, N Katz, P G Kühl, and W Rascher.
    • Department of Pediatrics, Friedrich-Alexander-University of Erlangen, Nürnberg, Germany.
    • Crit. Care Med. 1999 Jun 1;27(6):1191-4.

    ObjectiveDuring nitric oxide inhalation, methemoglobinemia needs to be monitored. We compared six commercially available instruments and one manual method for methemoglobin measurements. In addition, we studied whether and to what degree methylene blue interferes with methemoglobin measurements.DesignIn vitro methodologic study.SettingResearch laboratory in a university hospital.PatientsFive healthy volunteers from whom red blood cells were obtained.InterventionsMethemoglobinemia was generated in a red blood cell suspension by nitric oxide; methemoglobin was measured with six commercial instruments and one manual photometric method to calculate variation coefficients and to determine the differences between the devices. Methemoglobin was measured with and without the addition of methylene blue with two instruments. Measurements were performed immediately after the addition of methylene blue.Measurements And Main ResultsAll six commercially available instruments had variation coefficients of <0.1 at methemoglobin concentrations of 5%, whereas the manual photometric method did not reach a variation coefficient of <0.1 at 8% of methemoglobin. Apart from two devices that measured slightly but significantly higher methemoglobin levels, all instruments measured similar values of methemoglobin when the same samples were determined simultaneously. Higher concentrations of methylene blue (10, 40, 100 microM) reduced substantially the apparent concentrations of methemoglobin. Interference by methylene blue was most pronounced at low methemoglobin levels.ConclusionsWith some limitations, all commercial instruments that were tested performed adequately for the monitoring of methemoglobinemia. Methylene blue interferes with the methemoglobin measurements in a dose-dependent manner.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…