Intensive care medicine · Nov 1997
Randomized Controlled Trial Comparative Study Clinical TrialRespiratory comfort of automatic tube compensation and inspiratory pressure support in conscious humans.
- J Guttmann, H Bernhard, G Mols, A Benzing, P Hofmann, C Haberthür, D Zappe, B Fabry, and K Geiger.
- Section of Experimental Anaesthesiology, University of Freiburg, Germany.
- Intensive Care Med. 1997 Nov 1; 23 (11): 111911241119-24.
ObjectiveTo compare the new mode of ventilatory support, which we call automatic tube compensation (ATC), with inspiratory pressure support (IPS) with respect to perception of respiratory comfort. ATC unloads the resistance of the endotracheal tube (ETT) in inspiration by increasing the airway pressure, and in expiration by decreasing the airway pressure according to the non-linear pressure-flow relationship of the ETT.DesignProspective randomized single blind cross-over study.SettingLaboratory of the Section of Experimental Anaesthesiology (Clinic of Anaesthesiology; University of Freiburg).SubjectsTen healthy volunteers.InterventionsThe subjects breathed spontaneously through an ETT of 7.5 mm i.d. Three different ventilatory modes, each with a PEEP of 5 cmH2O, were presented in random order using the Dräger Evita 2 ventilator with prototype software: (1) IPS (10 cmH2O, 1 s ramp), (2) inspiratory ATC (ATC-in), (3) inspiratory and expiratory ATC (ATC-in-ex).Measurements And Main ResultsImmediately following a mode transition, the volunteers answered with a hand sign to show how they perceived the new mode compared with the preceding mode in terms of gain or loss in subjective respiratory comfort: "better", "unchanged" or "worse". Inspiration and expiration were investigated separately analyzing 60 mode transitions each. Flow rates were continuously measured. The transition from IPS to either type of ATC was perceived positively, i.e. as increased comfort, whereas the opposite transition from ATC to IPS was perceived negatively, i.e. as decreased comfort. The transition from ATC-in to ATC-in-ex was perceived positively whereas the opposite mode transition was perceived negatively in expiration only. Tidal volume was 1220 +/- 404 ml during IPS and 1017 +/- 362 ml during ATC. The inspiratory peak flow rate was 959 +/- 78 ml/s during IPS and 1048 +/- 197 ml/s during ATC.ConclusionsATC provides an increase in respiratory comfort compared with IPS. The predominant cause for respiratory discomfort in the IPS mode seems to be lung over-inflation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.