Comparative Study
Transcranial optical spectroscopy--A comparison of the TOS 96 and INVOS 3100 cerebral oximeters.
- G Litscher, I Hadolt, and E Eger.
- Department of Anesthesiology and Intensive Care Medicine, University of Graz, Austria.
- Biomed Tech (Berl). 1998 May 1;43(5):133-6.
AbstractWe evaluated two user-friendly, commercially available transcranial cerebral oximeters (TOS 96, Tostec, Tokyo, Japan and INVOS 3100, Somanetics, Troy, USA) for use in adults. We compared the absolute values (%) of regional cerebral oxygen saturation (rSO2) from 108 measurements in 54 healthy volunteers (31 women, mean age 35.8 +/- 16.3 years and 23 men, mean age 35.4 +/- 9.2 years). The mean rSO2 values obtained by the TOS 96 and INVOS 3100 systems were 72.3 +/- 8.4 % and 72.6 +/- 7.2% respectively. Significant differences (p < 0.01) were found between females and males with the INVOS equipment. The study shows that anatomical (thickness of the skull) and technical (calibration) aspects have to be taken into consideration when interpreting absolute rSO2 values.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.