- Jakub Godzik, Samuel Kalb, Eduardo Martinez-Del-Campo, Anna G U S Newcomb, Vaneet Singh, Corey T Walker, Steve W Chang, Brian P Kelly, and Neil R Crawford.
- *Department of Neurosurgery, Barrow Neurological Institute †Spinal Biomechanics Laboratory, Department of Neurosurgery Research, St. Joseph's Hospital and Medical Center, Phoenix, AZ ‡Medtronic, Inc., Memphis, TN.
- Spine. 2016 Aug 1; 41 (15): E902-7.
Study DesignHuman cadaveric biomechanical study.ObjectiveThe aim of this study was to evaluate the biomechanics of lumbar motion segments instrumented with the CD HORIZON Spire Z plate system (Spire Z), a posterior supplemental fixation spinous process plate, alone and with additional fixation systems.Summary Of Background DataPlates and pedicle screw/rod and facet screw implants are adjuncts to fusion. The plate limits motion, improving segmental stability and the fusion microenvironment. However, the degree to which the plate contributes to overall stability when used alone or in conjunction with additional instrumentation has not been described.MethodsStandard nondestructive flexibility tests were performed in 7 L2-L5 human cadaveric spines. Spinal stability was determined as mean range of motion (ROM) in flexion/extension, lateral bending, and axial rotation. Paired comparisons were made between five conditions: (1) intact/control; (2) Spire Z; (3) Spire Z with unilateral pedicle screw/rod system (Spire Z+UPS); (4) Spire Z with unilateral facet screw system (Spire Z+UFS); and (5) Spire Z with bilateral facet screw system (Spire Z+BFS). Stiffness and ROM data were compared using one-way analysis of variance, followed by repeated-measures Holm-Šidák tests.ResultsSpire Z was most effective in limiting flexion (20% of normal) and extension (24% of normal), but less effective in reducing lateral bending and axial rotation. In lateral bending, Spire Z+BFS and Spire Z+UPS constructs were not significantly different and demonstrated greater ROM reduction compared with Spire Z+UFS and Spire Z (P < 0.001). Spire Z+BFS demonstrated greatest stiffness in axial rotation compared with Spire Z+UPS (P = 0.025), Spire Z+UFS (P = 0.001), and Spire Z (P < 0.001). Spire Z+UPS was not significantly different from Spire Z+UFS (P = 0.21), and superior to Spire Z (P = 0.013).ConclusionThe Spire Z spinous process plate provides excellent immediate fixation, particularly for flexion and extension. While the hybrid Spire Z+BFS screw construct afforded the greatest stability, Spire Z+UPS demonstrated considerable promise.Level Of EvidenceN/A.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.