• Cardiology in the young · Aug 2016

    Accuracy of the Masimo SET® LNCS neo peripheral pulse oximeter in cyanotic congenital heart disease.

    • Michael J Griksaitis, Gemma E Scrimgeour, John V Pappachan, and Andrew J Baldock.
    • Paediatric Intensive Care Unit,Southampton Children's Hospital,University Hospital Southampton,United Kingdom.
    • Cardiol Young. 2016 Aug 1; 26 (6): 1183-6.

    UnlabelledIntroduction Non-invasive peripheral pulse oximeters are routinely used to measure oxyhaemoglobin saturation (SpO2) in cyanotic congenital heart disease. These probes are calibrated in healthy adult volunteers between arterial saturations of ~75 and 100%, using the gold standard of co-oximetry on arterial blood samples. There are little data to attest their accuracy in cyanotic congenital heart disease. Aims We aimed to assess the accuracy of a commonly used probe in children with cyanotic congenital heart disease.MethodsChildren with cyanotic congenital heart disease admitted to the Paediatric Intensive Care Unit with an arterial line in situ were included to our study. Prospective simultaneous recordings of SpO2, measured by the Masimo SET® LNCS Neo peripheral probe, and co-oximeter saturations (SaO2) measured by arterial blood gas analysis were recorded.ResultsA total of 527 paired measurements of SpO2 and SaO2 (using an ABL800 FLEX analyser) in 25 children were obtained. The mean bias of the pulse oximeter for all SaO2 readings was +4.7±13.8%. The wide standard deviation indicates poor precision. This mean bias increased to +7.0±13.7% at SaO2 recordings <75%. The accuracy root mean square of the recordings was 3.30% across all saturation levels, and this increased to 4.98% at SaO2 <75%.ConclusionsThe performance of the Masimo SET® LNCS Neo pulse oximeter is poor when arterial oxyhaemoglobin saturations are below 75%. It tends to overestimate saturations in children with cyanotic congenital heart disease. This may have serious implications for clinical decisions.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…