-
Intensive care medicine · May 1995
Case ReportsMisleading information from pulse oximetry and the usefulness of continuous blood gas monitoring in a post cardiac surgery patient.
- J P Benson, B Venkatesh, and V Patla.
- University Department of Anaesthesia, Queen Elizabeth Hospital, Edgbaston, Birmingham, UK.
- Intensive Care Med. 1995 May 1;21(5):437-9.
AbstractWe report a case where misleading readings form a pulse oximeter were compared with the results obtained from a continuous multiparameter blood gas sensor (Paratrend 7 - Biomedical Sensors UK). The continuous mode of recording was useful in determining the right clinical course for the patient and serves to show how pulse oximetry may demonstrate misleading results in low flow states.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.