Cardiac output by transesophageal echocardiography using continuous-wave Doppler across the aortic valve.
- P L Darmon, Z Hillel, A Mogtader, B Mindich, and D Thys.
- Department of Anesthesiology, St. Luke's-Roosevelt Hospital Center, College of Physicians and Surgeons, Columbia University, New York, New York 10025.
- Anesthesiology. 1994 Apr 1;80(4):796-805; discussion 25A.
BackgroundThe use of transesophageal echocardiography for the determination of cardiac output (CO) has been limited to date. We assessed the capability of aortic continuous-wave Doppler transesophageal echocardiography to determine CO (DCO) in a transgastric long-axis imaging plane of the heart by comparing DCO to thermodilution CO (TCO).MethodsDCO was determined in 63 consecutive patients undergoing cardiac surgery. Aortic valve area was obtained from the transverse short-axis view of the valve assuming a triangular shape for the valve orifice. Stroke volume was calculated as the product of velocity-time integral and aortic valve area: stroke volume = velocity-time integral x aortic valve area. DCO was calculated off-line, by multiplying stroke volume with heart rate: DCO = stroke volume x heart rate.ResultsThe aortic valve orifice was easily imaged in all patients. Excellent-quality continuous-wave Doppler flow profiles were obtained in nearly all (62 of 63). A total of 109 DCO determinations were performed. Mean DCO was 4.35 +/- 1.18 l.min-1 (range 2.02-7.42 l.min-1), and mean TCO was 4.41 +/- 1.17 l.min-1 (range 2.24-8.94 l.min-1). Very high correlation and agreement were found between the two methods: DCO = 0.94 x TCO + 0.19, r = 0.94, SEE (standard error of the estimate) = 0.41 l.min-1; 95% confidence interval = 0.06 +/- 0.83 l.min-1. Relative changes from pre- to postbypass CO (delta) also showed a strong correlation (delta DCO = 0.93 x delta TCO + 5.4%, r = 0.82, SEE = 17.8%). For CO changes greater than 10%, Doppler was in accordance with thermodilution in 43 of 45 measurements. DCO repeatability coefficient was 0.51 l.min-1.ConclusionsCompared to thermodilution, continuous-wave Doppler measurements of blood flow velocity across the aortic valve in the transesophageal echocardiographic transgastric view allow accurate CO determination.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.