• Resuscitation · Mar 2014

    Comparative Study

    Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT(®) point-of-care analyser.

    • Esther S Veldhoen, Karen M K de Vooght, Martijn G Slieker, Anne B Versluys, and Nigel McB Turner.
    • Paediatric Intensive Care Unit, UMC Utrecht, Wilhelmina Children's Hospital, The Netherlands. Electronic address: E.S.Veldhoen@umcutrecht.nl.
    • Resuscitation. 2014 Mar 1; 85 (3): 359-63.

    BackgroundIntraosseous access is used in emergency medicine as an alternative when intravenous access is difficult to obtain. Intraosseous samples can be used for laboratory testing to guide treatment. Many laboratories are reluctant to analyse intraosseous samples, as they frequently block conventional laboratory equipment. We aimed to evaluate the feasibility and accuracy of analysis of intraosseous samples using an i-STAT(®) point-of-care analyser.MethodsIntravenous and intraosseous samples of twenty children presenting for scheduled diagnostic bone marrow aspiration were analysed using an i-STAT(®) point-of-care analyser. Sample types were compared using Bland Altman plots and by calculating intraclass correlation coefficients and coefficients of variance.ResultsThe handheld i-STAT(®)point-of-care analyser proved suitable for analysing intraosseous samples without technical difficulties. Differences between venous and intraosseous samples were clinically acceptable for pH, base excess, sodium, ionised calcium and glucose in these haemodynamically stable patients. The intraclass correlation coefficient was excellent (>0.8) for comparison of intraosseous and intravenous base excess, and moderate (around 0.6) for bicarbonate, sodium and glucose. The coefficient of variance of intraosseous samples was smaller than that of venous samples for most variables.ConclusionAnalysis of intraosseous samples with a bedside, single-use cartridge-based analyser is feasible and avoids the problem of bone marrow contents damaging conventional laboratory equipment. In an emergency situation point-of-care analysis of intraosseous aspirates may be a useful guide to treatment.Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.